15 Inch Patient Monitor M15

Model:           M15

Utruemed brings you the advanced 15-Inch Patient Monitor M15. We have designed this product to meet the evolving needs of healthcare professionals. The 15”color TFT LCD screen is the biggest USP of this product which also features a human voice alarm to handle emergencies during patient care. The standard configuration of this product includes ECG,SpO2,NIBP,TEMP,Resp, and PR. It can also be used ac2-IBP, C.O, EtCO2 and Thermal Recorder. Healthcare professionals have access to real-time visualization of vital signs and critical parameters, enabling informed decisions and improving the level of care. Our M15 Monitor is ideal for hospitals, clinics, and other healthcare settings.

Additional information

Product Details:

Product Features:

  •  15”color TFT LCD screen.
  •  Optional touch screen.
  •  Unique human voice alarm.
  • 720 hours trend data and 12 hours of all waveforms.
  •  Wire and wireless network.
  •  14 languages include: English,French,German,Spanish,Portuguese,Russian and etc.

 

Configuration:

  •  5-lead ECG,SpO2,NIBP,TEMP,Resp,PR,Li-ion battery.

Options:

  •  2-IBP, C.O, EtCO2; Thermal Recorder.

 

Technical Specification
ECG
Lead model 5 Leads (I, II, III, AVR, AVL,AVF, V)
Gain: 2.5mm/mV, 5.0mm/mV, 10mm/mV,20mm/mV
Heart Rate  
Measure range Adult:15-300 bpm, Neonatal/Pediatric: 15-350 bpm
Accuracy 1%
Resolution 1 bpm
Sensitivity > 200 μVP-P
Differential Input Impedance > 5 MΩ
CMRR Diagnostic Mode>90 dB, Monitor Mode>110 dB, Surgery, Mode>110 dB
Offset Potential ± 300mV
Leakage Current < 10 μA
Pace Pulse Detect Range:± 2~± 700mV, Width:0.1~2ms, Rise Time:10-100 μs
Pace Pulse Rejection Range:± 2~± 700mV, Width:0.1~2ms, Rise Time:10-100 μs
Baseline Recovery < 3 s After defibrillation.
ECG Signal Range ± 5 mV (Vp-p)
Bandwidth Surgery: 1-20 Hz, Monitor: 0.5-40 Hz, Diagnostic: 0.05-130 Hz
Calibration Signal: 1 mV p-p, ± 5% accuracy
ST measurement: Range:-2.0-+2.0 mV
Accuracy -0.8mV-+0.8mV,± 0.02mVor± 10%, which is greater;Other Range: unspecified
Arrhythmia 13 Types: ASYSTOLE, VFIB/VTAC, COUTPLET, BIGMINY,TRIGEMINY, R ON T, VT >2, BRADY, MISSED BEATS, PNP, PNC
RESP
Method Impedance between RA-LL or RA-LA
Differential Input Impedance >2.5 ΜΩ
Respiration Impedance Range 0.3-3Ω
Base Impedance Range 200 Ω-2000Ω
Bandwidth 0.3-2 Hz
Gain x0.25, x0.500, x1, x2, x4
Respiration Rate
Measurement Range Adult:0 ~120 BrPM,Neonatal / Pediatric: 0~150 BrPM
Resolution 1 BrPM
Accuracy 0-6 BrPM: unspecified, 7-150 BrPM: ± 2 BrPM or ± 2% use the greater.
Inspiriting current < 300 μA RMS max
Apnea Alarm 10 -40 s
NIBP
Method Oscillometry
Measure Mode Manual, Auto, STAT
Measure Interval in AUTO Mode 1,2,3,4,5,10,15,30,60,90,120,180,240,480 min
Measure Period in STAT Mode 5 min
Pulse Rate Range 40 ~240 bpm
Measure and Alarm Range
Adult Mode SYS:40~280 mmHg, DIA:10~220 mmHg, MEAN:20~240 mmHg
Pediatric Mode SYS:40~220 mmHg, DIA:10~160 mmHg, MEAN:20~170 mmHg
Neonatal Mode SYS:35~135 mmHg, DIA:10~100 mmHg, MEAN:20 ~110 mmHg
Static pressure accuracy ± 3mmHg
Resolution 1mmHg
Accuracy Maximum Mean Error: ± 5mmHg, Maximum Standard Deviation: 8mmHg
Overpressure Protection Adult:300 mmHg, Pediatric:240 mmHg, Neonatal:150 mmHg
SpO2
Measurement Range 0~100%
Resolution 1%
Accuracy 70% ~100% ± 2 %, 0%~69%: unspecified
Pulse Rate Measure and Alarm Range 20-250bpm
Resolution 1bpm
Accuracy ± 3bpm
TEMP
Channel 2
Measure and Alarm Range 0~50 ℃
Resolution 0.1℃
Accuracy (no sensor): ± 0.1℃(25℃ – 45℃), ± 0.2℃(other);

(include sensor): ±0.2℃(32℃ – 42℃) ,± 0.3℃ (other)