Model: M15
Utruemed brings you the advanced 15-Inch Patient Monitor M15. We have designed this product to meet the evolving needs of healthcare professionals. The 15”color TFT LCD screen is the biggest USP of this product which also features a human voice alarm to handle emergencies during patient care. The standard configuration of this product includes ECG,SpO2,NIBP,TEMP,Resp, and PR. It can also be used ac2-IBP, C.O, EtCO2 and Thermal Recorder. Healthcare professionals have access to real-time visualization of vital signs and critical parameters, enabling informed decisions and improving the level of care. Our M15 Monitor is ideal for hospitals, clinics, and other healthcare settings.
Product Features:
Configuration:
Options:
Technical Specification | |
ECG | |
Lead model | 5 Leads (I, II, III, AVR, AVL,AVF, V) |
Gain: | 2.5mm/mV, 5.0mm/mV, 10mm/mV,20mm/mV |
Heart Rate | |
Measure range | Adult:15-300 bpm, Neonatal/Pediatric: 15-350 bpm |
Accuracy | 1% |
Resolution | 1 bpm |
Sensitivity | > 200 μVP-P |
Differential Input Impedance | > 5 MΩ |
CMRR | Diagnostic Mode>90 dB, Monitor Mode>110 dB, Surgery, Mode>110 dB |
Offset Potential | ± 300mV |
Leakage Current | < 10 μA |
Pace Pulse Detect | Range:± 2~± 700mV, Width:0.1~2ms, Rise Time:10-100 μs |
Pace Pulse Rejection | Range:± 2~± 700mV, Width:0.1~2ms, Rise Time:10-100 μs |
Baseline Recovery | < 3 s After defibrillation. |
ECG Signal Range | ± 5 mV (Vp-p) |
Bandwidth | Surgery: 1-20 Hz, Monitor: 0.5-40 Hz, Diagnostic: 0.05-130 Hz |
Calibration Signal: | 1 mV p-p, ± 5% accuracy |
ST measurement: | Range:-2.0-+2.0 mV |
Accuracy | -0.8mV-+0.8mV,± 0.02mVor± 10%, which is greater;Other Range: unspecified |
Arrhythmia | 13 Types: ASYSTOLE, VFIB/VTAC, COUTPLET, BIGMINY,TRIGEMINY, R ON T, VT >2, BRADY, MISSED BEATS, PNP, PNC |
RESP | |
Method | Impedance between RA-LL or RA-LA |
Differential Input Impedance | >2.5 ΜΩ |
Respiration Impedance Range | 0.3-3Ω |
Base Impedance Range | 200 Ω-2000Ω |
Bandwidth | 0.3-2 Hz |
Gain | x0.25, x0.500, x1, x2, x4 |
Respiration Rate | |
Measurement Range | Adult:0 ~120 BrPM,Neonatal / Pediatric: 0~150 BrPM |
Resolution | 1 BrPM |
Accuracy | 0-6 BrPM: unspecified, 7-150 BrPM: ± 2 BrPM or ± 2% use the greater. |
Inspiriting current | < 300 μA RMS max |
Apnea Alarm | 10 -40 s |
NIBP | |
Method | Oscillometry |
Measure Mode | Manual, Auto, STAT |
Measure Interval in AUTO Mode | 1,2,3,4,5,10,15,30,60,90,120,180,240,480 min |
Measure Period in STAT Mode | 5 min |
Pulse Rate Range | 40 ~240 bpm |
Measure and Alarm Range | |
Adult Mode | SYS:40~280 mmHg, DIA:10~220 mmHg, MEAN:20~240 mmHg |
Pediatric Mode | SYS:40~220 mmHg, DIA:10~160 mmHg, MEAN:20~170 mmHg |
Neonatal Mode | SYS:35~135 mmHg, DIA:10~100 mmHg, MEAN:20 ~110 mmHg |
Static pressure accuracy | ± 3mmHg |
Resolution | 1mmHg |
Accuracy | Maximum Mean Error: ± 5mmHg, Maximum Standard Deviation: 8mmHg |
Overpressure Protection | Adult:300 mmHg, Pediatric:240 mmHg, Neonatal:150 mmHg |
SpO2 | |
Measurement Range | 0~100% |
Resolution | 1% |
Accuracy | 70% ~100% ± 2 %, 0%~69%: unspecified |
Pulse Rate Measure and Alarm Range | 20-250bpm |
Resolution | 1bpm |
Accuracy | ± 3bpm |
TEMP | |
Channel | 2 |
Measure and Alarm Range | 0~50 ℃ |
Resolution | 0.1℃ |
Accuracy | (no sensor): ± 0.1℃(25℃ – 45℃), ± 0.2℃(other);
(include sensor): ±0.2℃(32℃ – 42℃) ,± 0.3℃ (other) |